REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 5 - Maintenance
COSMED srl
•
Aromatic/halogenated hydrocarbons.
Never clean parts with metal brushes, steel wool or other scrubbing tools. All parts must not be exposed to
temperatures above 60 °C (140 °F).
Reprocessing Water Characteristics
Sterile water rinse is recommended when available, although tap potable water is also acceptable for use in
cleaning COSMED components.
User Responsibility
•
Carefully follow these instructions and the instructions provided with the reprocessing solutions.
•
Observe any relevant local legal provisions, as well as any organizational hygiene requirements, in addition
to these instructions (e.g.: use of gloves, protective masks, etc…).
•
Assign trained and competent individuals for the handling and reprocessing of the components.
•
DO NOT submerge the device or pour cleaning liquids over or into the device.
To minimize the potential for infection and ensure optimal performance, make sure that:
•
Device and re-processable parts are cleaned and disinfected prior the very first use.
•
Disposable items are disposed of after every test.
•
Parts showing defects on the material surfaces (cracks, brittleness) are disposed of.
•
Standard personal protective (PPE) measures (hand hygiene, gloves, gown, mask, face shields) are taken
as necessary.
•
Re-processable parts are reprocessed immediately after use.
Be sure that solutions for disinfection are:
•
Applied thoroughly to all surfaces of the items to be disinfected.
•
Used according to instructions and contact times recommended by the manufacturer.
•
Always stored and prepared in clearly marked containers to prevent accidental use/misuse.
Symbols
Immersion cleaning/disinfection
Wiping cleaning/disinfection
Disposable and Reusable Items for each Test Mode
The following table summarizes the disposable items and the ones requiring cleaning procedures for each
measurement mode.
Test Mode
Disposables
Reusable(cleaning required)
All
Device surfaces
Summary of Contents for Q-NRG
Page 1: ...Q NRG User Manual...
Page 2: ......
Page 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Page 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Page 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 151: ......