REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 5 - Maintenance
COSMED srl
CALIBRATION
Calibration Syringe 3
liters for Resting/Low
Flows
REF C00600-02-11
Syringe Small
adapter
REF
C04833-01-20
Corrugated tube for
adapter
REF A-108-300-012
Reprocessing Methods for Reusable Items
The following tables summarize the required post-test reprocessing procedures for each reusable items.
Device case
Level of
Reprocessing
Low-level disinfection
When
Immediately after use
Pre-treatment
Close all protection caps to correspondent receptacles. Wipe off any visible soil with a moist sponge
or towel
Manual Cleaning
When cleaning device w/o turning it off be sure to lock the touchscreen touching and holding the
lock button. Wipe surface with a soft cloth sprayed with cleaning solution (e.g. Oxivir® Tb RTU).
Dispose of the soft cloth
Disinfection
Repeat the manual cleaning. Wipe device surface with another soft cloth moistened with water.
Dispose the soft cloth.
Drying
Air dry
Material Facts
ABS plastic enclosure, glass display surface, Silicone rubber button and caps
Summary of Contents for Q-NRG
Page 1: ...Q NRG User Manual...
Page 2: ......
Page 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Page 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Page 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 151: ......