REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 3 - Q-NRG Operation
COSMED srl
B
A
A
Dead space affecting patient safety
B
Dead space affecting results accuracy
CANOPY Mode
•
The veil must be carefully wrapped around the patient to avoid air leaks which may cause VO2 and VCO2
underestimation.
•
Blower flow rate within the selected averaging interval shall be constant.
MASK Mode
•
VO2 and VCO2 accuracy relies on the measurement of the entire exhaled volume. Positioning and securing
the mask must ensure that there are no leak.