REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 5 - Maintenance
COSMED srl
Overview
To keep Q-NRG clean, safe and efficient, it must be properly maintained as outlined in this chapter.
•
Any service operations not specified in this manual should only be performed by authorized personnel only
in accordance with the service manual.
•
All parts (except single use items) are not shipped cleaned and should be re-processed prior to first use
according to the related instructions in this section.
•
All materials used in the construction of device and its accessories are non-toxic and pose no safety risks to
the patient or user.
•
The device should be turned off with the power cable and calibration cylinder disconnected prior to
cleaning, disinfecting, maintaining, servicing and/or inspecting the device.
•
The device must not be serviced or maintained while in use with a patient.
•
Inspect all re-processable parts after each phase of cleaning and disinfection (pre-treatment, cleaning,
and disinfection). Closely examine for signs of corrosion, damaged surfaces, split-offs, discolorations, and
impurities and remove damaged parts.
•
If, after cleaning and disinfection, parts remain dirty, repeat cleaning and disinfection procedures.
•
Because some environments cause a quicker collection of lint and dust than others, inspect the fan filter
and replace it more often when needed.
•
Dispose of all parts removed from the device according to your institution’s protocol. Follow all local,
state, and federal regulations with respect to environmental protection, especially when disposing of the
electronic device or parts of it (for example, oxygen cell, pneumatic cartridge, batteries).
Recommended Maintenance Schedule
Procedure
When?
Performed by
Note
Cleaning and disinfection
Before first use and after
every use
Expert
Refer to instructions
Preventive Maintenance
Every 350 tests
Expert
Replacement of:
- internal pneumatic cartridge
- cooling fan filter
O2 sensor replacement
Within 1 month after
device alert
Expert
Expected lifespan in room air is 12
months. Higher O2% may reduce
lifespan.
Firmware update
If available
Expert
Check www.cosmed.com for updates
Software update
If available
Expert
Check www.cosmed.com for updates
and refer to OMNIA user manual
Battery removal
Transportation of installed
Li-Ion batteries is
prohibited
Expert
Refer to www.iata.org for travelling
rules related to Li-ion batteries
Battery replacement
If necessary
Expert
Refer to instructions
Summary of Contents for Q-NRG
Page 1: ...Q NRG User Manual...
Page 2: ......
Page 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Page 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Page 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 151: ......