Operating Q-NRG in CANOPY Mode
1. Real-time data are displayed on-screen, measurements are displayed after approximately 90 seconds.
2. Data are stored automatically after 2 minutes (or as defined in the Settings/Advanced), tap
Start Recording
to store data immediately. Use this time to adjust dilution flow in order to achieve a CO2% that remains
between 0.5% and 1.5% (best compromise to avoid patient hyperventilation while maintaining accuracy).
Note:
Hyperventilation usually occurs during the first few minutes of the test; avoid adjusting dilution flow
until patient is comfortably relaxed. Measured values immediately after dilution flow changes are not
reliable; Average interval should not include dilution flow adjustments.
During test execution the user interacts with the device via the following commands:
Command
Function
Start Recording
Begin storage of data, test time will reset
Stop Recording
End test storing all data
Info
Review/edit additional test info (UN, Fasting Period, Resting Period, Test Position,
Body Temperature, Agitation Level, Sedation Level, Test Purpose and Operator
Notes)
MARKER
Insert a marker
Abort
End test discarding all data
/-
Adjust blower flow (6 available rates)
Actual
Shows real-time values (updated every 30 seconds)
AVG
Shows averaged results within the specified interval (manually adjustable using
the touch display)
Data
“Data” tab shows Realtime values in tabular format (updated every 30 seconds)
Summary of Contents for Q-NRG
Page 1: ...Q NRG User Manual...
Page 2: ......
Page 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Page 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Page 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 151: ......