REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 5 - Maintenance
COSMED srl
External Turbine Flowmeter
Level of Reprocessing
High-level disinfection
When
Immediately after use
Pre-treatment
Rinse in water at 22-40°C (72-104°F)
Manual Cleaning
Soak for 3 minutes in an enzymatic solution (e.g. Cidezyme) at room temperature (>22°C, 72°F).
Using a soft bristle brush, scrub the item submerged in the cleaning solution during the 3 min-
utes. Wipe while rinsing for 5 minutes in water. Inspect all surfaces and item features to ensure
that it is visibly clean, repeat Manual Cleaning if not.
Disinfection
Soak the component in disinfectant (e.g. Cidex OPA Solution) at room temperature (>22°C, 72°F)
for 12 minutes. Rinse three (3) times in a volume of room temperature (>22°C, 72°F) water, large
enough to completely submerge the part, for at least 3-minutes for each rinsing cycle.
Drying
Dry the parts immediately after the disinfection steps by using filtered air (oil-free, low-germ
and low-particle) to blow them off/out or dab with a lint free cloth and let air dry. Inspect all
surfaces and item features to ensure that they are visibly dry, repeat Drying if not.
Note
Do not place the turbine under running water. Do not use Accelerated Hydrogen Peroxide (AHP)
disinfectants to disinfect COSMED turbines since they can be chemically attacked, and the device
may not perform well.
Material Facts
ABS plastic enclosure (black), MABS (Methyl Methacrylate / ABS) clear plastic enclosure, Fan
blades (PET film), stainless steel axes, jewels, PC conveyors.
Calibration Syringe
Level of
Reprocessing
Low-level disinfection
When
Immediately after use
Pre-treatment
Wipe soil with a moist sponge or towel
Manual Cleaning
Wipe device surface with a soft cloth sprayed with cleaning solution (e.g. Oxyvir® Tb RTU). Dispose
the soft cloth.
Disinfection
Repeat manual cleaning. Wipe device surface with a soft cloth moistened with water. Dispose the
soft cloth.
Drying
Air dry
Note
Do not use any cleaning products or chemicals on the inside of the syringe. Dry thoroughly before
use. Device requires recalibration every year, date on syringe.
Material Facts
Cylinder: anodized aluminum, Seal: nitrile rubber, Piston rod: aluminum, End plate: POM
Summary of Contents for Q-NRG
Page 1: ...Q NRG User Manual...
Page 2: ......
Page 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Page 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Page 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Page 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Page 151: ......