Parameters of datalogger inputs
U0110M, U0110G
________________________________
Measured value:
Internal temperature
Temperature range:
(-
20 to +60) °C
Accuracy:
± 0.4 °C
(*)
(*)
During battery charging temporary measurement
inaccuracy can occur.
Response time:
t
63
< 6 min, t
90
< 15 min (temperature step
20 °C,
air circulation approx. 1 m/s)
Display resolution:
0.1 °C
Recommended calibration interval:
2 years
U0121M, U0121G
________________________________
Measured values:
2 x temperature sensed by an external COMET
Pt1000/E series probe
Temperature range:
(-20
0 to +260) °C, Pt1000/3850 ppm sensor
Measuring current: approx. 0.5 mA with pulses of
approx. 60 ms length
Input accuracy (without probes):
±0.2 °C in a range of (-200 to +100) °C
±0.2 % of the measurement value in a range of (+100
to +260)
°C
The accuracy of the device with a temperature probe
connected to it is determined by the above input
accuracy and the accuracy of the probe employed.
Connection:
Two-wire
connection
with
the
possibility
of compensation of the error caused by the cable
resistance. The probe is provided with a 3-pin M8 ELKA
3008V connector. Its connection is shown in
The recommended length of the Pt1000/E probe cable
is maximum 15 m, a length of 30 m must not be
exceeded. Shielded cables are recommended.
Summary of Contents for U0110G
Page 5: ...IE LGR UxxxxM 23 5 Dataloggers UxxxxM and UxxxxG U0141M U0141G models...
Page 61: ...IE LGR UxxxxM 23 61 Dimensions U0110M U0110G U8410M U8410G U0121M U0121G...
Page 62: ...62 IE LGR UxxxxM 23 U0141M U0141G U3120M U3120G U4440M U4440G...
Page 63: ...IE LGR UxxxxM 23 63 U3121M U3121G U3631M U3631G...
Page 64: ...64 IE LGR UxxxxM 23 U0141TM U0141TG U0843M U0843G U5841M U5841G U6841M U6841G U7841M U7841G...