IE-LGR-UxxxxM-23
69
Annex 6: SMS message format
SMS message format for SYSTEM ALARM:
The message header includes:
SYSTEM ALARM<enter>
„Description of the device DescDev“<enter>“
Depending on the type of alarm, the corresponding text:
Data memory limit overflow,
Battery warning,
Battery empty,
Battery Ok,
External power lost,
External power Ok;
Date/Time Error,
Date/Time Ok;
Measuring Error on input: ….;
Measuring Error END on input: ….,
Configuration error,
Configuration Ok,
The time of the event, such as „20170808 10:44“
Texts are fixed, user-immutable, if multiple events occur simultaneously
and the length of the message exceeds 160 characters (max message length),
then they will be divided into multiple messages.
Format of alarm SMS message from measured value:
Description of the device DescDev .... max. 32 chars <enter>
Name of the measured location: max. 16 chars <enter>
Te
xt „Actual value:
measured value + unit (or binary input status)<enter>
Text ¨ALARM¨
or
(Text „ALARM END“) <enter>
Description of the alarm: max. 16 chars
Separator comma + gap
Start time / end time of the alarm, such as 2016/05/08 10:44 <enter>
Text „ALARM“ (Text „ALARM END“)
In the case of other alarms followed by <enter> and continues with information
about the next alarm
Description of the alarm:
max. 16 chars
Separator comma + gap
2 chars
Start time of the alarm, such as 2016/05/08 10:44
: :
: :
The length of one message is limited to 160 characters, if there is a requirement
to send more characters, the characters are divided into more messages.
Summary of Contents for U0110G
Page 5: ...IE LGR UxxxxM 23 5 Dataloggers UxxxxM and UxxxxG U0141M U0141G models...
Page 61: ...IE LGR UxxxxM 23 61 Dimensions U0110M U0110G U8410M U8410G U0121M U0121G...
Page 62: ...62 IE LGR UxxxxM 23 U0141M U0141G U3120M U3120G U4440M U4440G...
Page 63: ...IE LGR UxxxxM 23 63 U3121M U3121G U3631M U3631G...
Page 64: ...64 IE LGR UxxxxM 23 U0141TM U0141TG U0843M U0843G U5841M U5841G U6841M U6841G U7841M U7841G...