IE-LGR-UxxxxM-23
55
(*1)
During battery charging temporary measurement
inaccuracy can occur.
(*2)
Important information is provided in the chapter
U5841M, U5841G
________________________________
Measured values:
3 x voltage (0
– 10) V
1 x binary input. This input can be configured to voltage
or a dry contact (with open collector transistor
acceptance). Unlike the U0843M (U0843G) model, this
input can capture connected/disconnected voltage
status.
Range and input levels:
Voltage inputs:
Range: (0 to 10) V DC
Accuracy: ± 10 mV
Input resistance: approx. 130 k
Ω
Binary input configured for voltage signal:
Input voltage for the „L“ level: < 0.8 V
(*)
Input voltage for the „H“ level: > 2 V
Minimum applicable voltage: 0 V
Maximum applicable voltage: +30 V DC
(*)
When the input is disconnected, the device will
measure the „L“ level.
Binary input configured for a dry contact or an open-
collector transistor:
Contact resistance for the „switched-on“ state:
< 10 kΩ
Contact resistance for the "switched-off" state:
> 2 MΩ
Excitation voltage: approx. 3 V
Minimum state duration necessary for latching
the state: 1s
Connection:
Pluggable terminal block, maximum wire cross section
1.5 mm
2
. The maximum cable length is 30 m. It is
recommended to use shielded cables.
CAUTION: The inputs are not galvanically isolated!
Recommended calibration interval:
2 years
Summary of Contents for U0110G
Page 5: ...IE LGR UxxxxM 23 5 Dataloggers UxxxxM and UxxxxG U0141M U0141G models...
Page 61: ...IE LGR UxxxxM 23 61 Dimensions U0110M U0110G U8410M U8410G U0121M U0121G...
Page 62: ...62 IE LGR UxxxxM 23 U0141M U0141G U3120M U3120G U4440M U4440G...
Page 63: ...IE LGR UxxxxM 23 63 U3121M U3121G U3631M U3631G...
Page 64: ...64 IE LGR UxxxxM 23 U0141TM U0141TG U0843M U0843G U5841M U5841G U6841M U6841G U7841M U7841G...