6- 14
6-1-10 Lever Ratio (LEVER)
Parameter name
Setting value
Remarks
Lever ratio (LEVER)
If a measurement point is on a lever, the measurement value can
be displayed as detected value
×
lever ratio.
0.1 to 100.0
[Default: 1.0]
The figure below shows a lever ratio of 2:1.
When the lever ratio is set to "2," the judgment value will be twice a normal measurement value.
Operation procedure
Note
If the lever ratio is set to a value that is out of range, digital display MAIN (white) shows ######.
The display then returns to the lever ratio setting screen, allowing the lever ratio to be set again.
■
■
2 秒長押し
/
選択
キャンセル
桁移動
数値変更
確定
HI. SE T
LO . S E T
HY ST E R
T E ACH
T O L <±>
ADVANC
CALC
CO PY
BANK
PR. VAL
PR. O BJ
PR. SAVE
DIRE CT
SPE E D
O UT PUT
MAINT E
CALI B
RE SE T
Long press
for 2 s
Select
Commit
Cancel
Shift between
digits
Change numerical
values
2
:
1