5-6 Bank Mode
In this mode, HIGH and LOW setting values can be written into designated banks (1 to 3) and read out
from those banks.
By using the bank function, settings for items to be measured can be stored in the bank in advance,
and these settings can then be read out easily when required.
Setting items written into and read out from banks
All setting items
HIGH and LOW setting values [Initial state]
HIGH and LOW setting values, and presets
Methods of writing into/reading from banks
Using keys
Through external input (for reading out only)
If Bank A or Bank B input is set to external input, the external input can be used to read out the settings
stored in the banks.
Each bank is read out by selecting a particular combination of ON/OFF inputs into Bank A and Bank B
inputs.
Ensure that the bank input is of 20 ms duration or longer.
The combinations of Bank A and B inputs are shown in the following table.
Bank A
Bank B
Bank 1
ON
OFF
Bank 2
OFF
ON
Bank 3
ON
ON
For reference
Because the setting values written into and read out from the banks are stored in internal
memory, they are retained after the power is turned off. (Values can be re-written approximately
one million times.)
Related information
The setting items written into and read out from the banks can be changed in setting mode.
For details of how to set them, refer to "6-5 Bank Settings."
For details of how to set Bank A or B input to external input, refer to "6-2-3 External Input."
■
■
■
■