6- 39
6-2-8 Alarm Settings
Parameter name
Setting value
Remarks
Alarm delay time (DELAY)
Sets the delay time (changeable in 1 ms increments) from the
time the sensor head comes into upthrust position to the time the
alarm activates.
1 to 1,000
[Default: 1,000]
Upthrust check
(PRS.CHK)
Sets the upthrust check ON/OFF.
ON / OFF
[Default: ON]
Upthrust check setting value
(PRS.SET)
Sets the threshold level of the upthrust check.
−199.9999 to 199.9999 (mm)
[Default: 10.0000]
Catching check
(CAT.CHK)
Sets the catching check ON/OFF.
ON / OFF
[Default: OFF]
Number of connected units check
(PRS.CHK)
If controllers are connected, this performs a check when the
power is turned ON confirming that the number of connected units
has not changed.
ON / OFF
[Default: OFF]
Upthrust
"Upthrust" indicates a state in which the spindle has been pushed in too much. Upthrust can cause damage to
the sensor head. If the upthrust check is set to ON, an alarm is output before upthrust occurs, protecting the
sensor head from damage.