![China Electronics Technology Instruments AV3672 Series User Manual Download Page 72](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790072.webp)
3 Introduction to Use
3.5 Data Analysis
68
3.5.4.1 Creating and Editing of Ripple Limit Line
The ripple limit test can be done to several traces at the same time. The ripple limit can be set after the trace is
activated.
1) Setting method
a) Menu path:
[Analysis] > [Test] > [Ripple test]
. The
test
dialog box will appear.
b) Click the
[Table]
button in the dialog box to open the ripple limit table.
Fig. 3.62 Ripple Test Dialog Box
2) Ripple limit table
Type
Start stimulus
Stop stimulus
Maximum fluctuation
1
ON
100.000kHz
2.000GHz
5.000dB
2
ON
2.000GHz
3.000GHz
10.000dB
3
OFF
100.000kHz
3.000GHz
100.000dB
Fig. 3.63 Ripple Limit Table
“Type” box
The “Type” box is used to set the limit test type:
ON: enable the test of the limit segment.
OFF: disable the test of the limit segment.
“Start stimulus” box and “Stop stimulus” box
Set the start and stop stimulus value of the ripple limit segment.
“Maximum fluctuation” box
Set the maximum fluctuation of the test line of the ripple limit segment test.
3.5.4.2 Setup of Ripple Test
After creating the ripple limit line, select the mode of displaying or hiding the ripple limit line of one trace. Even if
the limit line is hided, the corresponding ripple test is still effective. You can select to send the warning sound and
display the failure sign (FAIL) in case of the failure of ripple test. When the ripple test function is enabled, the trace
of the failure part is red in the default mode. You can change the
limit failure color
in the “
Color setting
” of the
right-click menu of the mouse, and the trace color of the passing part remains unchanged.
Menu path:
[Analysis] > [Test] - [Ripple test
]. The
test
dialog box will appear.
The ripple test can only be done at the actual sweep measurement point. The instrument can still pass the ripple test
if only a few sweep points are applied and the performance indicators of the DUT may not conform to the
requirements. Therefore, sufficient sweep points should be applied in actual measurement of the limit test.
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...