10 Fault Diagnoses and Repair
10.3 Error Information
297
521
Prompt
The calibration type does not match the selected measurement type. The
correction function is shut down or correction is not allowed.
Some kinds of calibration are only applicable to some S-parameters. For example, the
Port 1 calibration for S11 is not applicable to Port 1 calibration of S22.
522
Prompt
More standard kits should be measured.
524
Prompt
Not full double-port calibration.
525
Prompt
Two calibration kits are used.
526
Prompt
No user calibration data suitable for the channel are found.
Correction cannot be done as a result of no calibration.
527
Prompt
The standard calibration kit is not required for the calibration type.
This error is caused if one undesired standard calibration kit is measured during
calibration.
528
Prompt
The electrical calibration system cannot be configured. Check whether the
module is connected properly..
Before electrical calibration, normal communication between the network analyzer
and electrical calibration module should be established at first, and the module should
be confirmed; otherwise, electrical calibration cannot be done.
529
Prompt
Out-of-limit data: beyond the design range.
Key parameters include the sweep type, start frequency, frequency span and point
number.
530
Prompt
Execution error: the data dimensions do not match.
531
Prompt
Execution error: the array cleared.
532
Prompt
Execution error: the array function is invalid.
533
Prompt
Execution error: the array subscript is wrong.
534
Prompt
Execution error: the array subscript is out of limit.
535
Prompt
Execution error: the matrix rank is wrong.
536
Prompt
Execution error: CPU.
537
Prompt
Execution error: the module cannot be deleted.
538
Prompt
Execution error: the module cannot be written.
539
Prompt
Execution error: no input.
540
Prompt
Execution error: the system is busy and the command is invalid.
541
Prompt
Electrical calibration: the module is not on the RF channel. Connect the module
to the specified port.
The RF channel is not connected to the electrical calibration module, and no signal is
found in the corresponding calibration measurement.
542
Prompt
Execution error: no space for new calibration kit. Delete one calibration kit.
Key parameters include the sweep type, start frequency, frequency span and point
number.
543
Prompt
Execution error: the space is not enough.
544
Prompt
Execution error: other matrix errors.
545
Prompt
Execution error: unequal rank.
546
Prompt
Execution error: few constant ranks.
547
Prompt
Execution error: few variable ranks.
548
Prompt
Perform pre-calibration.
549
Prompt
Execution error: bug in the dynamic library of electrical calibration drivers or
invalid module number.
550
Prompt
Execution error: accidental error from the electrical calibration driver.
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...