7 Calibration
7.2 Selection of Calibration Type
211
4) Through type response and isolation
a) Calibration accuracy: medium.
b) Measurement parameter: S-parameter of transmission measurement.
c) Standard calibration kit: through type device and two load (one respectively for each port).
d) System error correction:
➢
Transmission tracking
➢
Crosstalk
e) Measurement application:
➢
Transmission measurement in any direction.
➢
Improvement of the dynamic range of the system by means of isolation calibration.
Attention
Isolation calibration
Isolation calibration is not allowed if all the ports are not respectively connected with one load at the same time.
5) One-port (reflection)
a) Calibration accuracy: high
b) Measurement parameter: S11, S22, S33 and S44.
c) Standard calibration kit: open-circuit device, short-circuit device and load.
d) System error correction:
➢
Direction
➢
Source matching
➢
Reflection tracking
e) Measurement application: reflection measurement of any port.
6) Full 2-port SOLT
a) Calibration accuracy: high
b) Measurement parameters: all
c) Standard calibration kit: open-circuit device, short-circuit device, load and through type device.
d) System error correction:
➢
Direction
➢
Source matching
➢
Reflection tracking
➢
Crosstalk
➢
Transmission tracking
e) Measurement application:
➢
Measurement of all S-parameters.
➢
Improvement of the measurement accuracy by 12 error corrections.
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...