11 Technical Indicators and Measurement Methods
11.2 Measurement Methods
318
first segment covers 10MHz
~
4GHz. Choose 7 test points (10MHz, 50MHz, 100MHz, 1GHz, 2GHz, 3GHz and
4GHz) within the band. Input the worst harmonic suppression value out of theses points in the Performance
Test Record. The second segment covers 4GHz
~
26.5GHz. Choose 6GHz, 8GHz, 10GHz, 12GHz, 14GHz,
16GHz, 18GHz, 20GHz, 22GHz and 24GHz, 10 points in total,
(
AV3672A only uses the first 4 points
)
. Input
the worst harmonic suppression value in the Record; AV3672C/D/E have three segments. The first one is the
same with that of AV3672A/B, the second part covers 4GHz
~
13.5GHz. Choose 6GHz, 8GHz, 10GHz and
12GHz as test points, input the worst harmonic suppression value in the Record; The third segments covers
13.5GHz
~
67GHz, choose 14GHz, 16GHz, 18GHz, 20GHz, 22GHz, 24GHz, 26GHz, 28GHz, 30GHz and 32GHz as
test points
(
AV3672C/D only uses the first 6 points
)
. Input the worst harmonic suppression value out of
the ten points in the Record;
5)Connect the RF cable to Port2/3/4, press
【
Meas
】
→
【
S22
】
/
【
S33
】
/
【
S44
】
, repeat Step
(
4
)~
(
5
)
, they should meet relevant performance requirements.
Fig. 11.3 Block Diagram of Test of Harmonic/Non-harmonic Suppression Test
11.2.4 Test of Maximum Output Power of Port
11.2.4.1 Items of Test of Maximum Output Power of Port
Test the maximum output power indicators of the port of AV3672 series vector network analyzers.
11.2.4.2 Required Instrument and Accessories
1)
Power meter
AV2434 or ML2437A
2)
Power probe
AV71711 or MA2475A
11.2.4.3 Steps of Test of Maximum Output Power of Port
1) As shown in Fig. 11.4, connect the system and preheat is for 60min at least.
2) Press [Reset] > [Power] > [+20] and the ENTER key.
3) Press [Sweep] > [Sweep Time] > [20] and the ENTER key.
4) Press [Frequency] > [Start frequency] and enter the start frequency 10MHz. Press [Stop frequency] and enter the
stop frequency 50MHz.
5) As shown in Fig. 11.4, calibrate/clear the power meter, connect the power probe to Port 1 and record the
minimum reading.
6) Successively change the sweep frequency of the AV3672 series vector network analyzer. In the high power
mode or filter mode required in the test, click [Channel] > [Hardware Configuration] > [Path Configuration]. Select
the high power mode or filtering mode for the low-frequency segment of Source 1. The test result is the maximum
output power of Port 1.
7) Connect the power probe to Port 2/3/4. Click [Meas] > [S22]/[S33]/[S44]. Repeat Step (4) to (6) and measure the
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...