![China Electronics Technology Instruments AV3672 Series User Manual Download Page 300](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790300.webp)
10 Fault Diagnoses and Repair
10.3 Error Information
296
10.3.2 Program Error Information
Error information formats
In the remote control mode, error information is recorded in the error log. Query the error information with the
command “SYSTem:ERRor?”. Error information formats are as follows:
"Time:"
“Error No.:”
“Error level:”
“Error type:”
Example:
Time: 20:40:21, 2014/11/16
Error No.: 1400
Error level: wrong
Error type: grammar error
Program error information is divided into two types:
➢
Negative error codes defined in the SCPI standard, not detailed here.
➢
Positive error codes of instrument characteristics, with details shown in the following table.
Table 10.2 List of Instrument Characteristic Error Information
Error code
Error Level Error illustration
512
Prompt
The secondary parameter (such as the power, intermediate frequency
bandwidth, sweep time and step mode) of the calibration state has been changed.
Type:
Changes of such parameters after calibration may result in inaccurate measurement.
513
Prompt
Calibration cannot be done as a result of failure to complete all standard
measurements.
This may be caused by enabling of the error correction function before all standard
measurements for calibration.
515
Prompt
The correction function cannot be enabled (no error correction data
corresponding to Channel <x> or Measurement <x>). Perform calibration.
No error correction data corresponding to the specified channel and measurement.
516
Prompt
The key parameters of the current instrument state do not match with the
calibration set, and the correction function is shut down.
Key parameters include the sweep type, start frequency, frequency span and point
number.
517
Prompt
The correction function is shut down as the interpolation function is shut down
and stimulus settings are changed.
The original stimulus settings should be applied for the most accurate calibration.
518
Prompt
Shut down the interpolation function, enable the correction function and recover
the previous stimulus settings.
519
Prompt
The stimulus range is beyond the original calibration settings, resulting in
shutdown of the correction function.
Correction data beyond the stimulus settings will not be generated.
520
Prompt
The calibration type of Channel <x> and Measurement <x> is “None”. Select the
calibration type by the [Cal] menu or with the keys.
Calibration will not be done if the calibration type is not selected or the invalid
calibration type is set. This error is caused by no calibration or calibration without
calibration type selection.
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...