![China Electronics Technology Instruments AV3672 Series User Manual Download Page 362](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790362.webp)
Appendix
Appendix 3 Frequency Offset Measurement
358
2) The independent intermediate frequency bandwidth and independent sweep time are valid for the main setting
and non-coupling receiver and invalid for the source.
3) The independent power is valid for the main setting and non-coupling source mode and invalid for the receiver.
3.3 Calibration of Frequency Offset Measurement
In the traditional calibration process, errors are corrected through the measurement reference channel. The stimulus
frequency is different from the response frequency in frequency offset measurement. Therefore, the traditional
standard calibration mode cannot be used for error correction, such as the full double-port SOLT mode. The source
and receiver power calibration mode should be applied for error correction in the frequency offset measurement.
Click
[Cal] > [Calibration...] > [Power calibration]
on the front panel and open the [Power calibration] option.
For details, refer to “7 Calibration”.
1) Source power calibration
➢
On the premise of no consideration to the receiver accuracy, set the accurate power corresponding to the
stimulus frequency.
➢
No channel copying function.
2) Receiver calibration
➢
Set a calibrated source.
➢
No channel copying function.
3) Mismatch error
If the connector of different impedance is connected between measurement points, mismatch error will occur. The
mismatch error can be eliminated by means of full double-port calibration in S-parameter measurement. This
problem cannot be solved easily in frequency offset measurement. The simplest way is to connect a high-quality
attenuator between the input end and the mixer output port. The ports can be matched effectively by adding one
attenuator between ports, so as to increase the attenuation value to be more than twice. The larger the attenuation
value is, the better the improvement effect is good, but the smaller the dynamic range of measurement is.
4) LO accuracy and stability
If the LO signal is not accurate or stable in the frequency offset measurement mode, the output signal will be
beyond the edge or band-pass of the intermediate frequency filter. At the same time, the LO power also plays an
important role in mixer measurement. Therefore, the LO power accuracy and stability must be ensured.
3.4 Typical Application of Frequency Offset Measurement
●
Frequency transformation loss (gain) measurement
·····························································
399
●
Frequency transformation compression measurement
··························································
400
●
Isolation measurement
·····························································································
400
●
Harmonic distortion measurement
················································································
402
●
Return loss and standing wave ratio measurement
······························································
402
3.4.1 Frequency Transformation Loss (Gain) Measurement
The frequency transformation loss is defined as the power ratio of the output frequency and input frequency at the
given LO power. The LO power has the influence on the frequency transformation loss. The frequency
transformation loss reflects the energy conversation efficiency of the input signal of the mixer from the current
frequency into the output frequency. If the frequency transformation loss is not flat within the measurement
frequency range, part of valuable information of the output signal will be lost.
Transmission measurement should be applied to measure the frequency transformation loss. The input signal power
will be measured by entering the specific stimulus and LO signal power. If the output frequency is different from
the input frequency, the frequency offset measurement must be applied.
Summary of Contents for AV3672 Series
Page 3: ......
Page 4: ...AV3672 Series Vector Network Analyzer Contents...
Page 5: ......
Page 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Page 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Page 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Page 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Page 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Page 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Page 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Page 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Page 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Page 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Page 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Page 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Page 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Page 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Page 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Page 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Page 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Page 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Page 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Page 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Page 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...