SKEMA 6
-
OPERATING INSTRUCTION
149
GB
149
GB
13. SKEMA 6 dental operatory disinfecting and maintenance plan
WHEN
PART
WHAT TO DO
SEE PARAGRAPH
Before starting work
Condensate drain cock
Drain condensate from air ducts
See paragraph 9.2
Spray ducts
(if the dental unit is turned off
for more than 48 hours)
Disinfect water ducts of sprays wih manual
cycle (if the S.S.S. system is present)
See paragraph 7.2.1
Disinfect water ducts of sprays wih semi-
automatic cycle (if the AUTOSTERIL system
is present)
See paragraph 7.2
Surgical suction fi lters
Place a VF CONTROL PLUS tablet inside
each fi lter
See paragraph 9.3
Single-use protective barriers
Put single-use barriers and fl exible protection
on dental chair and dental unit
/
Strumenti.
Lubricate micromotor and turbine handpie-
ces
See documentation enclosed
with the instrument
After each treatment
Spray ducts
Flush water ducts of sprays (if the TIME FLU-
SHING system is present)
See paragraph 7.6
Instruments
Sterilize or disinfect the outside
See documentation enclosed
with handpiece
Camera
Disinfect the outside
Do not use harsh products or acids
See paragraph 5.8
Surgical suction tubes
Perform an automatic fl ushing cycle (if present)
or draw approximately half a liter of
STER 3
PLUS
solution with each suction tube
See paragraph 9.4
Flexible protection
Sterilize the contaminated fl exible protection
/
Single-use barriers
Replace the single-use barriers
/
Contaminated surfaces
Disinfect with
STER 1 PLUS
/
At the end of the work day
Spray ducts
Disinfect water ducts of sprays wih manual
cycle (if the S.S.S. system is present)
See paragraph 7.2.1
Disinfect water ducts of sprays wih semi-
automatic cycle (if the AUTOSTERIL system
is present)
See paragraph 7.2
Bowl fi lter
Clean the fi lter in running water
The content must be disposed separately
See paragraph 7.1
Bowl.
Clean with commercial detergents designed
for ceramic materials.
Do not use harsh products or acids.
See paragraph 7.1
Surgical suction fi lter
Check the fi lter and replace it if the suction
capacity is reduced (code 97290068)
See paragraph 9.3
Surgical suction tubes.
Perform an automatic fl ushing cycle (if present)
or draw approximately half a liter of
STER 3
PLUS
solution with each suction tube
See paragraph 9.4
Hydraulic saliva ejector
Clean the saliva ejector fi lter
See paragraph 6.4
Flexible protection
Sterilize the fl exible protection present
/
Single-use barriers
Remove the single-use barriers used
/
Dental chair and dental unit
surfaces
Clean surfaces with
STER 1 PLUS
/