Page 9-10
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
LTV
®
SERIAL NUMBER: _____________
TEST
DESCRIPTION
STEP
#
MEAS.
VALUE
REQUIREMENT
PASS
(P) /
FAIL
(F)
Alarm Settings
Low
O
2
Pres alarm (LTV
®
1200 only)
4)B)
Set
O
2
% to 22% (LTV
®
1200 only)
Alarm must activate immediately
Low Minute Volume alarm
4)C)
LMV alarm set to 10
Alarm must activate at start of next
breath
Low
Pressure
alarm
4)D)
LP alarm set to 60 cmH
2
O
Alarm must activate at end of the next
inspiration
High
Pressure
alarm
4)E)
HP alarm 10 cmH
2
O below
monitored PIP
Alarm must activate on the next breath
O
2
Enrichment
(LTV
®
1200 only)
O
2
Inlet pressure @ 50 ± 2 PSIG,
%O
2
@ 60%
4)G)
External O
2
monitor = 55% to 65%.
No alarms activate.
Disc / Sense Alarm
High side sense line disconnected
from unit
4)I)
Alarm must activate at start of next
inspiration
Pressure Operation
(LTV 1200, 1150 only)
Control
Settings:
Pressure Assist / Control
Breath Rate = 10
Pressure Control = 45
Insp. Time = 1.5
PEEP = 5
Rise Time = As required
80
4)K)
Monitors should read as follows:
PIP:
46 to 54 cmH
2
O
No alarms activate.
Power Alarm
Return to normal ventilation mode
(LTV 1200, 1150 only).
Disconnect AC adapter from ventilator
4)L)
POWER LOST alarm activates.
Battery Level
LED illuminates showing
the charge level.
Ventilator continues to operate from
internal battery
Vent Inop Alarm
Clear the POWER LOST alarm. Turn
the ventilator off. Wait 15 seconds.
4)M)
After the ventilator is turned off, the
Vent
Inop
alarm sounds continuously for a
minimum duration of 15 seconds.
Vent Inop
LED illuminates continuously
for a minimum duration of 15 seconds.
Clear the Vent Inop alarm. Test is
complete.
4)N)
Confirming audible chirp must activate
after alarm is silenced
80
Set the Rise Time under the VENT OP menu as required (1 through 9, depending on test lung used), to
achieve 50 cmH
2
O of pressure.
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...