Page 7-4
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
Symptoms
Possible Causes
What to Do
A control doesn't
activate.
Set Value
knob
doesn't operate.
Control not active in
selected mode.
If a control is dimmed, it is not active in the
currently selected mode and changing its
setting does not affect ventilation. (See the
LTV
®
1200, 1150, or 1100 Ventilator
Operator's Manual for an explanation of
Bright, Dim and Blank Control
Displays
.)
Controls are locked.
If the controls are locked, a
LOCKED
message will be displayed when a control
is selected.
To unlock in
EASY
mode, press and
release the
Control Lock
button. To
unlock in
HARD
mode, press and hold the
Control Lock
button for 3 seconds. (See
the LTV
®
1200, 1150, or 1100 Ventilator
Operator's Manual for an explanation of
the
CTRL UNLOCK
feature and
Control
Lock
button.)
Control is not selected.
Before a control value can be changed, the
control must be selected. To select a
control, press the associated button. When
a control is selected, it is displayed at
normal intensity and all other controls are
dimmed. (See the LTV
®
1200, 1150, or
1100 Ventilator Operator's Manual for an
explanation of how to use the controls.)
Controls are limited.
A control's value may be limited by the
current settings of other controls. To
change the value of the current control,
change the value of the flashing controls.
(See the LTV
®
1200, 1150, or 1100
Ventilator Operator's Manual for an
explanation of
Control Limiting
.)
Wrong model selected.
If an incorrect model is selected, some
controls may not be lit and will not
operate. Verify the model number selected
in
VENT MTNCE
,
MODEL
matches the
model number on the front of the
ventilator. See
Chapter 6 -Maintenance &
Calibration
for instructions.
Continued...
A control doesn’t
activate
Set Value
knob
doesn’t operate
Front panel ribbon cable
not properly connected.
Remove the power board to access the
ribbon cable connection on the main
board. Disconnect and reconnect the front
panel ribbon cable connector. See
Main
Board Assembly
in Chapter 8 for
instructions. Handle the ribbon cable
carefully to avoid scratching or damaging
it.
Front panel ribbon cable
damaged.
Defective switch.
Replace the membrane switch panel and
overlay. See
Front Panel
in Chapter 8 for
instructions
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...