Page 9-26
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
12)
Perform a Flow Valve Leak for I/E Hold.
A) Power up ventilator in
VENT
MTNCE
mode (i.e.: using the
“three finger mode
90
” hold down
the
Control Lock
,
Manual
Breath
and
Select
buttons while
powering up the ventilator) and
enter
VENT MNTCE
.
B) Turn on
SERVO
mode.
C) Set the step position to 0 (in the
Tidal Volume
window)
D) Measure the flow at the Patient
Outlet port of the ventilator.
Flow must be less than 1 Lpm.
90
Using the 3-finger technique enables a modified mode of VENT MTNCE in which testing may be performed,
but Calibration procedures and changing Configuration settings can not.
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...