Page 7-34
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
Symptoms
Possible Causes
What to Do
CONTINUED …
Repeated
DISC/SENSE
alarms.
Exhalation drive line leaking
or loose.
Exhalation valve leaking
during inspiration.
Check the exhalation drive line at both the
ventilator and exhalation valve ends. Verify
the line is securely seated and not leaking.
Verify the exhalation valve is not leaking
during inspiration. If it is leaking, open the
exhalation valve and remove the
diaphragm. Reseat the diaphragm and
snap the valve cap back in place. See
Cleaning the Reusable Exhalation Valve
and Patient Circuit
in Chapter 4
for a
diagram of the exhalation valve assembly.
If necessary, replace the exhalation
diaphragm, or exhalation valve with a new
one.
Pressure Control or
Pressure Support set below
PEEP.
Verify these control values are
appropriately set.
Internal flexible tubing
occluded or pinched.
Internal flexible tubing has
pinhole leaks or leaking at
connections.
This problem will often be accompanied by
XDCR FAULT
alarms. Open the vent and
verify that none of the flexible tubes
connected to the solenoid manifold or
analog board are pinched or leaking. See
Internal Flexible Tubing
in Chapter 8 for
routing information.
Defective transducer on
analog board.
Replace the analog board. See
Analog
Board Assembly
in Chapter 8 for
instructions.
Defective auto zero
solenoid.
Replace the solenoid manifold. See
Solenoid Manifold Assembly
in Chapter 8
for instructions.
Repeated
XDCR
FAULT
alarms.
Internal flexible tubing
occluded or pinched.
Internal flexible tubing has
pinhole leaks or leaking at
connections.
This problem will often be accompanied by
XDCR FAULT
alarms. Open the vent and
verify that none of the flexible tubes
connected to the solenoid manifold or
analog board are pinched or leaking. See
Internal Flexible Tubing
in Chapter 8 for
routing information.
Defective transducer on
analog board.
Replace the analog board. See
Analog
Board Assembly
in Chapter 8 for
instructions.
Defective auto zero
solenoid.
Replace the solenoid manifold. See
Solenoid Manifold Assembly
in Chapter 8
for instructions.
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...