p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 9-29
LTV
®
SERIAL NUMBER: _____________
TEST
DESCRIPTION
STEP
#
MEAS.
VALUE
REQUIREMENT
PASS
(P) /
FAIL
(F)
Real-time Transducer Data
Airway Pressure (
AP
):
0
cmH
2
O to high flow XDCR port
6)A)
AP
display must read 0 ± 0.5 cmH
2
O
50
cmH
2
O to high flow XDCR port
6)B)
AP
display must read 50 ± 2 cmH
2
O
Flow Transducer Wide (
FDw
):
0
cmH
2
O to both flow XDCR ports
6)C)
FDw
display must read 0 ± 0.5 cmH
2
O
15
cmH
2
O to high flow XDCR port
6)D)
FDw
display must read 15 ± 0.5 cmH
2
O
15 cmH
2
O to high & low XDCR
ports
6)E)
FDw
display must read 0 ± 0.5 cmH
2
O
Flow Valve Differential Pressure:
At step position 500, turbine speed
10
6)G)
FVd
display must read 0 ± 0.5 cmH
2
O
At step position 100, turbine speed
4000
6)H)
FVd
display must read
≥
14 cmH
2
O
Oxygen Transducer (LTV
®
1200 only):
At 0 psig applied to O
2
inlet port
6)I)
% O
2
display must read 0 ± 0.5 psig
At 50 psig applied to O
2
inlet port
6)J)
%O
2
display must read 50 ± 2 psig
PEEP Pilot Pressure Transducer
0 cmH
2
O to all XDCR ports
6)K)
PPP display must read 0 ± 0.5
cmH
2
O
Patient Outlet Pressure
Measure Patient Outlet pressure at actual
turbine speed
7)E)
Must be
≥
55 cmH2O
Servo
Flow
@ 10,
Turbine
speed 4000,
% O
2
@ 30,
O
2
inlet pressure
40 psig
Press
Manual Breath
8)A)
Flow (BTPD) must be 10 Lpm ± 0.5 Lpm
%
O
2
must be 30 ± 3% (LTV
®
1200
only)
Flow
@ 10,
Turbine
speed 5000,
%O
2
@ 90, O
2
inlet pressure
50 psig
Press
Manual Breath
8)B)
Flow (BTPD) must be 10 Lpm ± 0.5
Lpm
%
O
2
must be 90 ± 5% (LTV
®
1200
only)
Flow
@ 50,
Turbine
speed 4000,
%O
2
@ 60, O
2
inlet pressure
60psig
Press
Manual Breath
8)C)
Flow (BTPD) must be 50 Lpm ± 7.5
Lpm
%
O
2
must be 60 ± 5% (LTV
®
1200
only)
Flow
@ 90,
Turbine
speed 7000,
%O
2
@ 90, O
2
inlet pressure
50 psig
Press
Manual Breath
8)D)
Flow (BTPD) must be 90 Lpm ± 13.5
Lpm
%
O
2
must be 90 ± 5% (LTV
®
1200
only)
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...