Page 7-30
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
Problem
Possible Causes
What To Do
The ventilator doesn’t
operate from external
power.
Defective AC source.
AC adapter power cord
loose.
Make sure the AC adapter is plugged into
a good AC source and is securely
connected to the ventilator. Verify the
power cord for the adapter is fully seated.
Defective AC adapter.
Replace the AC adapter.
Defective power board.
Replace the power board. See
Power
Board Assembly
in Chapter 8 for
instructions.
The ventilator does not
operate from internal
battery.
The ventilator shuts off
when external power is
removed.
Internal battery depleted.
If the internal battery is depleted, charge
the internal battery for 24 hours by
connecting the external AC adapter and
plugging it into a good AC source.
Internal battery not
connected.
If the internal battery is not connected, the
Charge Status
LED will show red.
Reconnect the internal battery. See
Internal Battery Pack
in Chapter 8 for
instructions.
Defective internal battery.
If the battery does not reach full capacity
after several charge cycles (vent should
run for more than 40 minutes on nominal
settings) replace the internal battery. See
Internal Battery Pack
in Chapter 8 for
instructions.
Defective power board.
If the battery won’t charge completely after
replacing it, there may be a problem with
the charge circuit on the power board.
Replace the power board. See
Power
Board Assembly
in Chapter 8 for
instructions.
Battery doesn’t reach
full charge.
Battery depletes too
quickly.
Internal battery deeply
discharged.
Charge the internal battery for 24 hours by
connecting the external AC adapter and
plugging it into a good AC source. If the
battery is deeply discharged, it may take
several cycles of charging and discharging
for the battery to reach a maximum charge.
Defective internal battery.
If the battery does not reach full capacity
after several charge cycles (vent should
run for more than 40 minutes on nominal
settings) replace the internal battery. See
Internal Battery Pack
in Chapter 8 for
instructions.
Defective power board.
If the battery won’t charge completely after
replacing it, there may be a problem with
the charge circuit on the power board.
Replace the power board. See
Power
Board Assembly
in Chapter 8 for
instructions.
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...