p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 7-3
Symptoms
Possible Causes
What to Do
Low Min Vol
display
flashing.
LOW MIN VOL
alarm
occurred.
The
Low Min Vol
display is flashed and
the
LOW MIN VOL
message is displayed
when a low minute volume alarm occurs.
The display will continue to flash even
after the condition clears. (See the LTV
®
1200, 1150, or 1100 Ventilator Operator's
Manual for an explanation of the
LOW
MIN VOL
alarm feature.)
O
2
% (O
2
Flush)
flashing.
LOW O2 PRES
or
HIGH
O2 PRES
alarm occurred.
The
O
2
% (O
2
Flush)
display is flashed and
the
LOW O2 PRES
or
HIGH O2 PRES
message is displayed when a low or high
O
2
pressure alarm occurs. The display will
continue to flash even after the condition
clears. (See the LTV
®
1200 Ventilator
Operator's Manual for an explanation of
the
LOW O2 PRES
and
HIGH O2 PRES
alarm features.)
Control display
flashing when setting
a control.
Control setting is limited.
A control's value can be limited by the
current settings of other controls. (See the
LTV
®
1200, 1150, or 1100 Ventilator
Operator's Manual for an explanation of
Control Limiting
.)
A display or LED does
not illuminate.
Wrong model selected.
If an incorrect model is selected, some
controls may not be lit and will not
operate. Verify the model number selected
in
VENT MTNCE
,
MODEL
matches the
model number on the front of the
ventilator. See
Chapter 6 -Maintenance &
Calibration
for instructions.
Misaligned LED.
If displays are operating but misaligned,
remove the main board and realign the
LEDs or displays. See
Main Board
Assembly
in Chapter 8
for instructions.
Defective LED or display.
Defective main board.
If a display is not operating during the
display test, replace the main board. See
Main Board Assembly
in Chapter 8 for
instructions.
Ventilator is running
but displays are
turned off.
Displays are blanked while
on battery power.
To conserve battery life while running from
the internal battery, most of the displays
are turned off when no changes are made
to the control settings for 60 seconds. To
turn the displays back on, touch any
control or button or turn the
Set Value
knob.
Defective main board.
Replace the main board. See
Main Board
Assembly
in Chapter 8 for instructions.
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...