Page 7-2
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
looking for under one section, you may find it listed under another. You may also be able to
diagnose the problem by reading sections with related symptoms. For information on
resolving problems that are not listed here, contact CareFusion using the contact
information at the front of this manual.
Displays and Buttons
Some of the symptoms listed in this section are part of the normal operation of the ventilator
and do not indicate a problem with the ventilator. They are included for information only.
Symptoms
Possible Causes
What to Do
Pressure Control
display flashing.
Pressure control breath
terminated by flow -
PC
FLOW TERM
is set to on.
Pressure control breaths are normally
terminated when the set inspiratory time
expires. Flow termination of pressure
control breaths is allowed when
PC FLOW
TERM
is set to ON (See the LTV
®
1200,
1150, or 1100 Ventilator Operator's
Manual for an explanation of this feature).
When a pressure control breath is
terminated by flow instead of time, the
Pres. Control
display is flashed.
Pressure Support
display flashing.
Pressure support breath
terminated by time - set
under
TIME TERM
.
Pressure support breaths are normally
terminated when the flow drops below the
set percentage of the peak flow. Pressure
support breaths may also terminate on
time when the variable time limit is
reached before the flow drops to the set
level. (See the LTV
®
1200, 1150, or 1100
Ventilator Operator's Manual for an
explanation of the
FLOW TERM
and
TIME
TERM
features). When a pressure support
breath is terminated based on time, the
Pres. Support
display is flashed.
High Pres Limit
display flashing.
HIGH PRES
alarm
occurred.
The
High Pres Limit
display is flashed
and the
HIGH PRES
message is
displayed when a high pressure alarm
occurs. The display will continue to flash
even after the condition clears. (See the
LTV
®
1200, 1150, or 1100 Ventilator
Operator's Manual for an explanation of
the
HIGH PRES
alarm feature.)
Check high and low pressure sense lines
to be sure they are correctly attached and
securely seated at both the ventilator and
wye ends.
Low Pressure
display flashing.
LOW PRES
alarm
occurred.
The
Low Pressure
display is flashed and
the
LOW PRES
message is displayed
when a low pressure alarm occurs. The
display will continue to flash even after the
condition clears. (See the LTV
®
1200,
1150, or 1100 Ventilator Operator's
Manual for an explanation of the
LOW
PRES
alarm feature.)
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...