Page 9-30
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J
LTV
®
SERIAL NUMBER: _____________
TEST
DESCRIPTION
STEP
#
MEAS.
VALUE
REQUIREMENT
PASS
(P) /
FAIL
(F)
Solenoids
Step
position 100,
turbine
speed 4000 9)A)
ExhPilot solenoid deactivated
(OFF):
Pressure at exhalation drive port must =
0 ± 0.1 cmH
2
O
ExhPilot solenoid activated (ON):
Pressure at exhalation drive port must
be
≥
14 cmH
2
O
Step
position 100,
turbine
speed 4000 9)B)
Purge solenoid deactivated (OFF):
Pressure at the High & Low flow
transducer ports = 0 cmH
2
O
Purge solenoid activated (ON):
Pressure at the High & Low flow
transducer ports
≥
14 cmH
2
O
PPin and PPout solenoids activated
(ON):
9)C)
Pressure at the exhalation drive port
must be
≥
14 cmH
2
O
Turbine
speed 10
9)D)
PPin solenoid deactivated (OFF):
Leakage rate must be
≤
1 cmH
2
O within
one minute
PPP value under the RT XDCR DATA
menu must be ± 0.5 cmH
2
O of the
measured pressure.
PPout solenoid deactivated (OFF):
9)E)
Pressure at exhalation drive port must =
0 ± 0.1 cmH
2
O
15
cmH
2
O to high flow transducer port: 9)F)
Apres solenoid activated (ON):
AP
: 0 cmH
2
O ± 0.5 after activation
FDw
: 0 cmH
2
O ± 0.5 after activation
15
cmH
2
O to low & high flow transducer
ports:
Pinch off inlet pressure, purge solenoid
activated for 3 sec., then deactivate.
9)G)
Calculated change in Pressure:
FDw
display must not change by more
than 0.5 cmH
2
O before and after
activating solenoid
15
cmH
2
O to low & high flow transducer
port:
9)H)
ExhDiffP
solenoid deactivated
(OFF):
FDw
display must read 0 ± 0.5 cmH
2
O
ExhDiffP
solenoid
activated (ON):
FDw
display must read AP ± 0.5 cmH
2
O
15
cmH
2
O to high flow transducer port: 9)I)
Pinch off the input pressure line
FDw
display must read 15 ± 0.5 cmH
2
O
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...