p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 9-25
11) Perform an LTM Compatibility test.
If an LTM Monitor is not available to perform this test, downloading the Event Trace onto a
computer using the Service Cable Assembly and Instructions contained in P/N 11485
(available seperatly or as part of the Maintenance and Calibration Kit, P/N 11566) will also
test for the proper function of the communication port.
A) Set the COM setting in Extended
Features to
MONITOR
B) Connect the LTM serial
communication cable to the UUT
COMM PORT
on the left side of
the UUT.
Ensure the patient circuit and test
lung(s) Are connected.
Observe the icon and red
exclamation mark for a minimum
of 30 seconds.
Display of small LTV icon appears on the
LTM. If the LTV
®
icon is not observed, or
shown only intermittently, then this
condition constitutes a FAIL for the LTM
compatibility test.
(LTM with software 1.17 or earlier) A red
exclamation mark is displayed to the
right of the LTV
®
icon. The exclamation
mark may be flashing.
C) Press the
Wave
button on the
LTM.
Select
the Wave screen.
Waveforms are displayed in all three
waveform charts.
The
Vte
value displayed to the left of the
lower waveform chart on the LTM must
be equal to the value shown in the LTV
®
status monitor window.
D) Set the
Low Minute Volume
alarm to 99 Lpm.
LOW MIN VOL
alarm will occur.
,
LTM status bar is flashing red with the
text “Low Minute Volume” displayed.
E) Reset the
Low Minute Volume
alarm by returning the
Low Min
Vol
to its previous value and then
pressing the
Silence Reset
button
twice on the LTV
®
.
Verify the red “Low Minute Volume”
status bar is removed from the LTM.
F) Disconnect the LTM serial
communication cable from the
LTV
®
COMM PORT
.
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...