p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 7-21
Symptoms
Possible Causes
What to Do
Condensation in sense
lines.
High or low sense lines are
occluded.
High or low sense ports in
the wye are occluded.
Verify lines are not occluded or pinched
and/or clear the lines with a low flow (less
than 10 Lpm) gas source.
Defective purge solenoids.
Replace the solenoid manifold. See
Solenoid Manifold Assembly
in Chapter 8
for instructions.
Ventilator is on, gas is
not delivered and
turbine is running.
Defective or disconnected
flow valve.
Defective turbine.
Using the Vent Maintenance mode
SERVO
test, determine if the turbine and flow valve
are working correctly by changing the
Turbine Speed, Flow, and Step Motor
Positions.
Verify all turbine and flow valve cables are
connected and all flexible tubes connected
to the analog board and solenoid manifold
are correctly and securely connected.
Perform the Advanced Vte Diagnostic
Procedures, see
page 7-23
.
Defective or disconnected
solenoids.
Using the Vent Maintenance mode
SOLENOID test, determine if all solenoids
are working correctly.
Perform the Advanced Vte Diagnostic
Procedures, see
page 7-23
.
Ventilator is on, gas is
not delivered and
turbine is not running.
RT XDCR DATA
item
TS
shows speeds at or
near 0.
Defective or disconnected
turbine.
The SERVO test may be used to set the
turbine speed explicitly. (See
Chapter 6 -
Maintenance & Calibration
for instructions
on using the servo test.) The turbine speed
can be viewed using the
RT XDCR DATA
display. (See
Chapter 3 - Real-Time
Transducer Data
for instructions on using
RT data.)
Verify all flow valve and turbine power
cables are connected. Verify all flexible
tubes connected to the analog board and
solenoid manifold are correctly and
securely connected.
Replace the turbine. See
Flow Valve
Assembly
or
Turbine Manifold
in Chapter 8
for instructions.
Defective motor board.
Replace the motor board. See
Motor
Board Assembly
in Chapter 8 for
instructions.
Defective power board.
Replace the power board. See
Power
Board Assembly
in Chapter 8 for
instructions.
Sense line connectors
wobble or loose.
Fitting connectors loosened
inside ventilator case.
Open the ventilator case and tighten the
nuts on the sense line and drive line
connectors. See
Back Panel
in Chapter 8
for instructions on opening the ventilator.
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...