p/n 18603-001, Rev. J
LTV
®
1200, 1150, and 1100 Ventilator Service Manual
Page 9-19
Insp Time
: 1.5 seconds
15% of average measured volume.
2)
Perform a Breath Rate test.
A) Set the ventilator as follows:
Tidal Volume
: 1500 ml
Breath Rate
: 10
Insp Time
: 3.0 seconds
B) Measure breath rate or the
period between the start of
two consecutive breaths.
Measured breath rate must be 10.0 ± 1.0 or the
interval must be 6.0 ± 0.5 sec.
3)
Perform a Pressure Control test (LTV 1200/1150 only).
A)
Set the ventilator as follows
:
Pressure Control
: 45 cmH
2
O and
Control
Mode
: Pressure
PEEP
: 5 cmH
2
O
Rise Time
: under the
VENT OP
menu as required (1 through 9,
depending on test lung used), to
achieve 50 cmH2O pressure
B) Measure the steady-state pressure
during inspiration (measured from 0
pressure baseline).
Measured steady-state pressure must be
46 to 54 cmH
2
O.
4)
Perform a Sensitivity test.
A) Set the ventilator as follows:
Sensitivity
: 2 Lpm
PEEP
: 5
Breath Rate
: 6
Tidal Volume
: 700
Inspiration Time
: 2.0
Control Mode
: Volume
B) Set
Leak Compensation
: OFF
Press and hold the
Select
button
Status monitor window displays
ALARM OP
Rotate
Set Value
knob
Status monitor window displays
VENT OP
Press
Select
button and rotate
Set
Value
knob
Status monitor window displays
LEAK
COMP
Press
Select
button and rotate
Set
Value
knob
Status monitor window displays
COMP OFF
Press
Select
button and press
Control Lock
button (2) times.
C) Observe the
Patient Effort
LED on
the upper right hand side of the
ventilator control panel while
creating a 4 Lpm leak past the
patient wye momentarily during the
exhalation portion of the respiratory
cycle.
The
Patient Effort
LED should momentarily
flash and an inspiration should immediately
begin.
If
the
Patient Effort
LED does not flash
and/or an inspiration is not given
immediately:
This condition constitutes a failure of the
sensitivity test.
D) Set
Leak Compensation:
ON
Summary of Contents for LTV 1100
Page 1: ...LTV 1200 1150 and 1100 Ventilator Service Manual P N 18603 001 Rev J ...
Page 15: ...Page 1 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 27: ...Page 2 12 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 39: ...Page 5 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 77: ...Page 6 38 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 121: ...Page 7 44 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 175: ...Page 8 54 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 307: ...Page A 14 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 313: ...Page B 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 316: ...p n 18603 001 Rev J LTV 1200 1150 and 1100 Ventilator Service Manual Page C 3 ...
Page 317: ...Page C 4 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 327: ...Page D 10 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 333: ...Page E 6 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...
Page 345: ...Page E 18 LTV 1200 1150 and 1100 Ventilator Service Manual p n 18603 001 Rev J ...