![Canon OMNERA 400A Operator'S Manual Download Page 97](http://html.mh-extra.com/html/canon/omnera-400a/omnera-400a_operators-manual_61478097.webp)
Operator´s Manual
0073-095-418-EN Rev. 5.1
OMNERA®
2020-09-25
Page 91
4.7.3 Pendulum Mode
4.7.3.1 General
The
Pendulum mode
is designed for non-vertical examinations of patients lying on the Table.
4.7.3.2 Flow of Operation
Select a
Pendulum mode
examination.
The System display will display the following, see Fig. 4-11
Fig. 4-11 Pendulum mode display
• The SID will be displayed.
Activate the mode by pressing the servo button.
• The OTC will automatically move to its programmed position.
• The OTC will move to the programmed SID.
Adjust the height of the Table stand.
• The OTC will follow the Table stand to maintain the programmed SID.
Press the arrow left or corresponding buttons on the table handle.
• Press the buttons for moving the OTC to the right or left, see item F and G at Fig. 3-41
. The tube move in the desired direction and the imaging unit move to stay
aligned with the tube.
4.7.3.3 Exposure
Exposure is possible when the OTC is not moving, and the servo button is activated.