2020-09-25
6.2 Collimator
WARNING!
Do not restart the System if inflammable liquids have leaked into the collimator.
CAUTION!
The collimator cover is not water resistant.
How to clean the collimator:
• Disconnect the whole System.
• Use non-abrasive cleaning products. Care must be taken to prevent liquid from entering
the collimator.