Operator´s Manual
Page 96
2020-09-25
4.8.2 NoWait Configuration
WARNING!
The Wallstand will immediately start tracking the height of the detector.
At the installation of the System, it is possible to select that the System shall not wait for the
user to move the Wallstand.
The flow will then be as follow;
Select a
Wall Flexible mode
examination.
• The SID will be displayed.
Activate the mode by pressing the servo button.
• All movements are available.
• The OTC will automatically move to its programmed position.
• The OTC will move down and align with the Wallstand detector.
• The OTC will follow the Wallstand detector to stay aligned.
• The OTC reaches the correct height and stops.
• The
SID
value is changed to the new distance (if selected distance towards the Wallstand
is changed).
4.8.2.1 Exposure
Exposure is possible when the OTC stands still, and the servo mode indication light is fixed.