4.5 Manual Mode
CAUTION!
Be aware that there is no tracking between the x-ray tube and the detector when the
System is in Manual mode. Therefore, take extra care that the x-ray beam is within the
active area of the detector before performing an exposure.
The System can be set to a
Manual mode
from the display. In the
Manual mode
it is possible
to move the System manually in the room and perform exposure. In
Manual mode
no track-
ing will be performed. The
Manual mode
can be selected when the System is in one of the
following modes:
•
WallFlexible
•
TableFlexible
•
FilmTracking
•
Pendulum table
4.5.1 Activation of Manual Mode
The servo state can be either
Automatic mode
or
Manual mode
. A pop-up window according
to Fig. 4-5
Servo state selection pop-up window
, will appear with the two alternatives. When
the System is in
Manual mode
all movements are allowed and exposure can be performed in
any position.
Fig. 4-5 Servo state selection pop-up window
The
Manual mode
is activated by pressing the
Servo state
, see Fig. 4-5
. If the System is in
TableFlexible
,
FilmTracking
or
Pendulum mode
the ta-
ble and the OTC will be shown without connection when
Manual mode
is selected. In the
same way, the Wallstand will be shown without connection to the OTC in
Manual mode
, see
Wallstand and Table shown without connection to the OTC in Manual mode
Fig. 4-6 Wallstand and Table shown without connection to the OTC in Manual mode