8.2 Ceiling Suspended X-ray Tube Support
8.2.1 General
Rotation range ceiling (beta)
>340°
Rotation range tube arm (alpha)
>±135°
Column (Z stroke)
1750 mm
8.2.2 Configuration
OTC
The OTC is a mechanical part of an X-ray
system.
8.2.3 Weight
OTC
Maximum 240 kg
Tube and collimator
40 kg maximum allowed weight
Ceiling wagon
95 kg
Column
40 kg
Ceiling rail Y (4 m standard)
28 kg
8.2.4 Electrical Characteristics
Mains voltage
230 VAC, 50/60 Hz center tapped single
phase 4 A
8.2.5 Classification
Classification according to IEC 60601-1.
Class
Class I equipment. All dead metal parts of
the equipment are electrical connected to
protective earth.
Applied part
Type B
Protection against ingress of water
IPXO
Mode of operation
Intermittent operation: 20%, maximum 1
min. ON / 4 min. OFF
Use of anaesthetic mixtures
The equipment is not suitable for use in the
presence of flammable anaesthetic mix-
tures with air or with oxygen or with nitrous
oxide.