2.8.4 Wall stand
Note!
The patient or operator is allowed to lean against the Wall stand, patient handle or armrest,
but
not
to put any weight on it.
2.8.4.1 Standard Version of the Wall stand
Possible squeezing hazard areas are indicated in Fig. 2-6
. Getting stuck in
the imaging unit slide opening (1), causes squeezing hazard, if the imaging unit is moving
downward or upward (Z-direction).
CAUTION!
If the motorized movement is operated in high speed level, it is not allowed to have
patient sitting or standing in the surroundings of the Wall stand.
Max 25kg/55 lbs
Z
1
Max 10 kg/22 lbs
Fig. 2-6 Warning labels
1. Slide opening of the imaging unit
2.8.4.2 Weight Restrictions
• The maximum weight to put on the wall stand lateral armrest is restricted to 25 kg/ 55 lbs.
• For the wall stand detector holder the maximum weight is set to 10 kg/ 22 lbs.
2.8.4.3 Indication of Power to Wall stand
The device is powered when the indicator light on the electrical box is lit.