![Canon OMNERA 400A Operator'S Manual Download Page 134](http://html.mh-extra.com/html/canon/omnera-400a/omnera-400a_operators-manual_61478134.webp)
OMNERA®
0073-095-418-EN Rev. 5.1
Operator´s Manual
Page 128
2020-09-25
8.3 Table 0055
8.3.1 Column
Lowest table top position (from the floor to
the table top surface)
550 mm
Column (Z stroke)
380 mm
8.3.2 Table Top
Table top dimension
2424 mm X 850 mm
Table top transparent area
2400 mm X 613 mm
Table top thickness
21.5 mm
Length of stroke, X direction
+/- 600 mm
Length of stroke, Y direction
+/- 150 mm
Movement range of the imaging unit
>650 mm
8.3.3 Weight
Table
Approximately 150 kg
Imaging unit
Approximately 21 kg
Table top
Approximately 47 kg
Maximum patient load
300 kg
8.3.4 Electrical Characteristics
Maximum power without external
electronics:
500 W
8.3.4.1 External Electrical Characteristics
The external electronics must be approved according to IEC60601-1.
If any external electronics is installed the end product must be tested according to
IEC60601-1.
Power output to external
110-240 VAC 50-60 Hz
Single phase 10 A
Power output external 24 VDC
24 VDC 3 A