![Canon OMNERA 400A Operator'S Manual Download Page 135](http://html.mh-extra.com/html/canon/omnera-400a/omnera-400a_operators-manual_61478135.webp)
Operator´s Manual
0073-095-418-EN Rev. 5.1
OMNERA®
2020-09-25
Page 129
8.3.5 Classification
Classification according to IEC 60601-1.
Class
Class I equipment.
All dead metal parts of the equipment are
electrical connected to protective earth.
Applied part
Type B
Protection against ingress of water
IPX2
Mode of operation
Intermediate use: 20% 1 min ON / 4 min
OFF
Use of anaesthetics mixtures
The equipment is not suitable for use in the
presence of flammable anaesthetics mix-
tures with air or with oxygen or with nitrous
oxide.
8.3.6 Attenuation Equivalent
Table top
< 0.9 mm AL at 3.7 mm HVL