4.8 Wallstand Examination Procedures
4.8.1 Wall Flexible Modes
4.8.1.1 General
The
Wall Flexible modes
are designed for examinations of patients standing up against a
vertical moving imaging unit.
The movement up/down of the OTC, will only change the tube position, the Wallstand will not
follow.
4.8.1.2 Wall Flexible Movements
There is 3 different Wall Flexible Modes, all accessible from the Arcoma Service program,
Adjust WallFlexible parameters
. The
Wall Flexible Movements
shall be set at the installation,
and will thereafter be valid for all Auto positions using the
Wall Flexible mode
.
Fig. 4-15 Adjust WallFlexible parameters.
•
Not Blocked
— All OTC movements are allowed.
•
Beta blocked and Sideways supervised
— Blocked beta movement, movement will turn
the servo off. It is possible to move a distance of half the detector length (landscape), be-
fore turning off the servo.
•
Beta and Sideways blocked —
It is only possible to move the OTC upwards and down-
wards. Blocked beta movement, movement will turn servo off. Blocked non-FFD direction-
al (X/Y) movement, movement will turn servo off and block exposure.