Guidance and manufacturers declaration - electromagnetic immunity
Voltage dips,
short interrup-
tions and voltage
variations on
power supply in-
put lines.
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the System requires
continued operation during power mains
interruptions, it is recommended that the
System be powered from an uninterrupted
power supply or battery.
Power frequency
(50/60 Hz) mag-
netic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environment.
Note!
U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturers declaration - electromagnetic immunity
The System is intended for use in the electromagnetic environment specified below. The customer or
the user of the System should assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the System, including cables, than the
recommended separation distance, calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
Conducted
RF IEC
61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80
MHz
d = 1,2 √p