4
Aesculap
®
S4 MIS instruments
Disassembling
The threaded tube FW734R and adjust-
ing screw FW707R
►
When using the adjusting screw FW707R with
S
4
percutaneous threaded tube FW734R: Screw the
adjusting screw
4
off the threaded tube FW734R,
see Fig.
A
.
Spacer FW709R
►
Remove the nut and the guide sleeve
a
from the
spacer
5
, see Fig.
B
.
Bar insertion instrument FW711R
►
Screw the inner part out of the bar insertion
instrument
6
, see Fig.
C
.
Distractor FW238R and distraction
blades, short FW706R
►
Activate slider
b
and remove the distraction
blades
7
from the distractor FW238R, see Fig.
D
.
►
Dismantle distractor FW238R with the distraction
blades.
WARNING
Excessive trocar insertion
depth in the event of a missing
end stop means a risk of
injury!
►
Use trocar FW271R only in
combination with the aim-
ing device FW258R.
WARNING
Delay in surgery due to the
outer sleeve slipping down
FW705R!
►
Secure the outer sleeves by
hand when inserting or
removing other instru-
ments (e.g. the thread cut-
ter, screwdriver).
■
The outer sleeve can be
reapplied using the posi-
tioning aid FW708R.
■
The outer sleeve can be
secured using the threaded
tube FW234R and the
adjusting screw, long
FW707R
CAUTION
Damage to the pedicle screws
and incorrect positioning due
to using the wrong screw-
driver!
►
Only use the guide instru-
ment FW708R for screw-
ing in monoaxial pedicle
screws.
CAUTION
Damage to the monoaxial ped-
icle screw when screwing in
with the FW708R with the
screwdriver in the wrong posi-
tion!
►
When screwing in the
monoaxial pedicle screw,
always insert the screw-
driver all the way into the
head of the pedicle screw.
Summary of Contents for Aesculap S4 MIS Spyder
Page 2: ...30 40 50 60 70 30 40 50 60 70 80 80 100 80 70 60 50 40 90 1 2 4 3 5 6 7 8...
Page 3: ...FW734R 4 7 b FW238R 5 a 6 A B C D...
Page 5: ...6 7 FW238R J K...
Page 119: ...115 S4 TA012384 S4 Spinal System S4 FRI S4 SRI S4 MIS S4 S4...
Page 122: ...118 Aesculap S4 MIS Aesculap https extranet bbraun com Aesculap 6 45 C...
Page 123: ...119 VAH FDA CE pH 8 www a k i org Rote Brosch re 6...
Page 124: ...120 Aesculap S4 MIS 55 C...
Page 125: ...121 FW705R FW709R FW711R FW712R FW714R FW716R FW734R 20 FW712R...
Page 126: ...122 Aesculap S4 MIS FW705R FW709R FW711R FW714R FW716R FW734R 20 FW705R FW709R FW714R FW716R...
Page 127: ...123 BBraun Stabimed T C F t I 15 2 pH 9 II 1 III 15 2 pH 9 IV 1 V...
Page 128: ...124 Aesculap S4 MIS I 15 1 5 II III 5 IV 5 V...
Page 130: ...126 Aesculap S4 MIS FDA CE BBraun Stabimed I 15 1 5 II T C F t I 15 2 pH 9 II 1...
Page 132: ...128 Aesculap S4 MIS Aesculap Aesculap STERILIT I JG600 STERILIT I JG598...
Page 191: ...187 S4 TA012384 S4 S4 FRI S4 SRI S4 MIS K S4 S4 X FW271R FW258R...
Page 194: ...190 Aesculap S4 MIS OP 6 45 C OP VAH FDA CE pH 8 www a k i org OP 6...
Page 195: ...191 55...
Page 198: ...194 Aesculap S4 MIS 1 15 1 5 II III 5 IV 5 V...