14
Aesculap
®
S4 MIS instruments
Inspection, maintenance and checks
►
Allow the product to cool down to room tempera-
ture.
►
After each complete cleaning, disinfecting and dry-
ing cycle, check that the product is dry, clean,
operational, and free of damage (e.g. broken insu-
lation or corroded, loose, bent, broken, cracked,
worn, or fractured components).
►
Dry the product if it is wet or damp.
►
Repeat cleaning and disinfection of products that
still show impurities or contamination.
►
Check that the product functions correctly.
►
Immediately put aside damaged or inoperative
products and send them to Aesculap Technical Ser-
vice, see Technical Service.
►
Assemble dismountable products, see Assembling.
►
Check for compatibility with associated products.
Packaging
►
Appropriately protect products with fine working
tips.
►
Store products with ratchet locks fully opened or
locked no further than in the first notch.
►
Close the locks so that the jaw tips just touch each
other.
►
Place the product in its holder or on a suitable tray.
Ensure that all cutting edges are protected.
►
Pack trays appropriately for the intended steriliza-
tion process (e.g. in Aesculap sterile containers).
►
Ensure that the packaging provides sufficient pro-
tection against recontamination of the product
during storage.
Steam sterilization
Note
FW711R: The product may only be sterilized in disas-
sembled condition.
Note
FW734R: the product can be sterilized either in disas-
sembled or in assembled condition.
Note
To avoid breakage due to stress crack corrosion, steril-
ize the product with the lock fully open or locked no fur-
ther than on the first ratchet tooth.
►
Check to ensure that the sterilizing agent will come
into contact with all external and internal surfaces
(e.g. by opening any valves and faucets).
►
Validated sterilization process
–
Disassemble the product
–
Steam sterilization through fractionated vac-
uum process
–
Steam sterilizer according to DIN EN 285 and
validated according to DIN EN ISO 17665
–
Sterilization using fractionated vacuum pro-
cess at 134 °C/holding time 5 min
►
When sterilizing several products at the same time
in a steam sterilizer, ensure that the maximum load
capacity of the steam sterilizer specified by the
manufacturer is not exceeded.
CAUTION
Damage (metal seizure/fric-
tion corrosion) to the product
caused by insufficient lubrica-
tion!
►
Prior to function checks,
lubricate moving parts (e.g.
joints, pusher components
and threaded rods) with
maintenance oil suitable
for the respective steriliza-
tion process (e.g. for steam
sterilization: Aesculap
STERILIT® I oil spray JG600
or STERILIT® I drip lubrica-
tor JG598).
Summary of Contents for Aesculap S4 MIS Spyder
Page 2: ...30 40 50 60 70 30 40 50 60 70 80 80 100 80 70 60 50 40 90 1 2 4 3 5 6 7 8...
Page 3: ...FW734R 4 7 b FW238R 5 a 6 A B C D...
Page 5: ...6 7 FW238R J K...
Page 119: ...115 S4 TA012384 S4 Spinal System S4 FRI S4 SRI S4 MIS S4 S4...
Page 122: ...118 Aesculap S4 MIS Aesculap https extranet bbraun com Aesculap 6 45 C...
Page 123: ...119 VAH FDA CE pH 8 www a k i org Rote Brosch re 6...
Page 124: ...120 Aesculap S4 MIS 55 C...
Page 125: ...121 FW705R FW709R FW711R FW712R FW714R FW716R FW734R 20 FW712R...
Page 126: ...122 Aesculap S4 MIS FW705R FW709R FW711R FW714R FW716R FW734R 20 FW705R FW709R FW714R FW716R...
Page 127: ...123 BBraun Stabimed T C F t I 15 2 pH 9 II 1 III 15 2 pH 9 IV 1 V...
Page 128: ...124 Aesculap S4 MIS I 15 1 5 II III 5 IV 5 V...
Page 130: ...126 Aesculap S4 MIS FDA CE BBraun Stabimed I 15 1 5 II T C F t I 15 2 pH 9 II 1...
Page 132: ...128 Aesculap S4 MIS Aesculap Aesculap STERILIT I JG600 STERILIT I JG598...
Page 191: ...187 S4 TA012384 S4 S4 FRI S4 SRI S4 MIS K S4 S4 X FW271R FW258R...
Page 194: ...190 Aesculap S4 MIS OP 6 45 C OP VAH FDA CE pH 8 www a k i org OP 6...
Page 195: ...191 55...
Page 198: ...194 Aesculap S4 MIS 1 15 1 5 II III 5 IV 5 V...