8
Aesculap
®
S4 MIS instruments
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning with
immersion disinfection
■
FW705R–FW709R
■
FW711R
■
FW712R
■
FW714R–FW716R
■
FW734R
■
Suitable cleaning brush
■
20 ml disposable syringe
■
Keep working ends open for clean-
ing.
■
When cleaning products with mov-
able hinges, ensure that these are in
an open position and, if applicable,
move the joint while cleaning.
■
Drying phase: Use a lint-free cloth
or medical compressed air
Chapter Manual cleaning/disinfection
and sub-chapter:
■
Chapter Manual cleaning with
immersion disinfection
Mechanical alkaline
cleaning and thermal
disinfection
■
FW712R
■
Place the product in a tray that is
suitable for cleaning (avoiding rins-
ing blind spots).
■
Connect components with lumens
and channels directly to the rinsing
port of the injector carriage.
■
Keep working ends open for clean-
ing.
■
Place products in the tray with their
hinges open.
Chapter Mechanical cleaning/disinfect-
ing and sub-chapter:
■
Chapter Mechanical alkaline clean-
ing and thermal disinfecting
Manual pre-cleaning
with brush and subse-
quent mechanical alka-
line cleaning and ther-
mal disinfection
■
FW705R–FW709R
■
FW711R
■
FW714R–FW716R
■
FW734R
■
Suitable cleaning brush
■
20 ml disposable syringe
■
Place the product in a tray that is
suitable for cleaning (avoiding rins-
ing blind spots).
■
FW705R/FW709R/FW714R–
FW716R: Connect components with
lumens and channels directly to the
rinsing port of the injector carriage.
■
Keep working ends open for clean-
ing.
■
Place products in the tray with their
hinges open.
Chapter Mechanical cleaning/disinfec-
tion with manual pre-cleaning and sub-
chapter:
■
Chapter Manual pre-cleaning with a
brush
■
Chapter Mechanical alkaline clean-
ing and thermal disinfecting
Summary of Contents for Aesculap S4 MIS Spyder
Page 2: ...30 40 50 60 70 30 40 50 60 70 80 80 100 80 70 60 50 40 90 1 2 4 3 5 6 7 8...
Page 3: ...FW734R 4 7 b FW238R 5 a 6 A B C D...
Page 5: ...6 7 FW238R J K...
Page 119: ...115 S4 TA012384 S4 Spinal System S4 FRI S4 SRI S4 MIS S4 S4...
Page 122: ...118 Aesculap S4 MIS Aesculap https extranet bbraun com Aesculap 6 45 C...
Page 123: ...119 VAH FDA CE pH 8 www a k i org Rote Brosch re 6...
Page 124: ...120 Aesculap S4 MIS 55 C...
Page 125: ...121 FW705R FW709R FW711R FW712R FW714R FW716R FW734R 20 FW712R...
Page 126: ...122 Aesculap S4 MIS FW705R FW709R FW711R FW714R FW716R FW734R 20 FW705R FW709R FW714R FW716R...
Page 127: ...123 BBraun Stabimed T C F t I 15 2 pH 9 II 1 III 15 2 pH 9 IV 1 V...
Page 128: ...124 Aesculap S4 MIS I 15 1 5 II III 5 IV 5 V...
Page 130: ...126 Aesculap S4 MIS FDA CE BBraun Stabimed I 15 1 5 II T C F t I 15 2 pH 9 II 1...
Page 132: ...128 Aesculap S4 MIS Aesculap Aesculap STERILIT I JG600 STERILIT I JG598...
Page 191: ...187 S4 TA012384 S4 S4 FRI S4 SRI S4 MIS K S4 S4 X FW271R FW258R...
Page 194: ...190 Aesculap S4 MIS OP 6 45 C OP VAH FDA CE pH 8 www a k i org OP 6...
Page 195: ...191 55...
Page 198: ...194 Aesculap S4 MIS 1 15 1 5 II III 5 IV 5 V...