Boonton 4540 Series RF Power Meter
Operation
4-50
4.5.29 Markers Menu (Modulated and Pulse Modes)
The Markers menu is used to configure measurement markers at specific points on the processed measurement waveform.
Automatic measurements may then be used to retrieve measurements at the two markers and in the interval between them.
Markers are used in Pulse Mode and Modulated Mode to perform measurements at or between two time offsets relative to the
trigger. The markers can only be placed on the visible portion of the trace as defined by the timebase and trigger delay
settings.
Main Pg2 > Markers > Marker 1 | 2
Description:
Set the time (x-axis-position) of marker 1 or 2 relative
to the trigger. Note that time markers must be
positioned within the time limits of the trace window
in the graph display. If a time outside of the display
limits is entered, the marker will be placed at the first
or last time position as appropriate.
Menu Type:
Numeric Value.
Range:
Current displayed timebase limits
SCPI Command:
MARKer[1|2]:POSItion:TIMe <numeric_value>
Main Pg2 > Markers > Delta Time
Description:
Displays the result of Marker 2 – Marker 1 in seconds.
This item is read only.
Menu Type:
Numeric value.
SCPI Command:
None.
Fig. 4-40 Markers Menu
(Modulated and Pulse)
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...