![Boonton 4541 Instruction Manual Download Page 59](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863059.webp)
Boonton 4540 Series RF Power Meter
Operation
4-11
4.5.7 Channel 1 | 2 Menu
The Channel menu contains items that affect a single measurement channel. The ―chann
el‖
is a full measurement path,
beginning with an external power sensor and sensor cable, and including data acquisition, processing of measurements, and
display of the processed information. The Model 4541 single channel power meter has only a Channel 1 menu, while the
dual-channel Model 4542 has separate color-coded menus for Channel 1 and Channel 2.
Main > Channel > Channel 1 | 2 > On/Off
Description:
Set or return the measurement state of the selected
channel. When ON, the channel performs
measurements; when OFF, the channel is disabled and
no measurements are performed.
Menu Type:
Toggle.
SCPI Command:
CALCulate[1|2]:STATe <Boolean>
Main > Channel > Channel 1 | 2 > Vert Scale
Description:
Set the power or voltage vertical sensitivity of the trace
display in channel units.
Menu Type:
Numeric Value.
Range:
Watts: 1 nW/Div to 50MW/Div
Volts: 1 mV/Div to 50kV/Div
dBm,
dBV,
dBr:
0.1 dB/Div to 50 dB/Div
SCPI Command:
DISPlay:TRACe[1|2]:VSCALe <numeric_value>
Main > Channel > Channel 1 | 2 > Vert Center
Description:
Set or return the power or voltage level of the
horizontal centerline of the graph for the specified
channel in channel units. If a change in the vertical
scale causes the center maximum value to be exceeded, the center will be forced to the
maximum value for the new range.
Menu Type:
Numeric Value.
Range:
Watts,
Volts: ±10,000 * Vertical Scale per divsion
dBm,
dBV,
dBr:
±200 dB
SCPI Command:
DISPlay:TRACe[1|2]:VCENTer <numeric_value>
Fig. 4-7 Channel 1|2 Menu
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...