Boonton 4540 Series RF Power Meter
Getting Started
3-15
STEP
PROCEDURE
1.
Select the
Calibration
menu soft key. Initiate autocalibration by selecting
AutoCal > Start
. The
AutoCal
routine
will calibrate the entire dynamic range of the sensor in approximately 1.5 minutes while reporting status via a dialog
box.
Figure 3-4 Autocalibration screen.
Note
You can
access the calibration menus directly using the
Cal/Zero
key. Pressing the key
repeatedly will cycle between each channel’s calibration menu. The calibration menus are
presented in channel order for all active channels. Active channels are defined as those that
are installed, turned on, and have a sensor connected. If there are not active channels,
pressing the
Cal/Zero
key will bring you to the Channel 1 calibration menu.
2.
If an error message appears after you have initiated
AutoCa
l, verify the following:
a. Is the sensor that is connected to the calibrator also connected to the channel indicated in the
Chan 1
>
Select
menu?
b. Are the sensor cable connections secure at both the sensor and instrument input channels?
c. Does the instrument pass its self-test (no errors reported when you pressed PWR)?
d. Are any attenuators or other devices that are connected between the sensor and the calibrator?
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...