![Boonton 4541 Instruction Manual Download Page 117](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863117.webp)
Boonton 4540 Series RF Power Meter
Operation
4-69
4.5.45 System > Calibrator Menu (Internal 50 MHz)
The Calibrator Menu is used to select and control the internal 50 MHz RF calibrator. It can also select the optional, external
1 GHz accessory calibrator (Model 2530). Note that this menu does not contain any items related to sensor calibration - it is
only for controlling the calibrator for use as a signal source. For sensor calibration information, refer to Section 3.5 and
Section 4.5.2 of this manual.
Main Pg2 > System > Calibrator > Cal Output
Description:
Enable or disable the RF output of the Calibrator.
Menu Type:
Multiple Choice.
Choices:
Off: Disabled
On: Provides 50 MHz CW output
SCPI Command:
OUTPut:INTernal:SIGNal <Boolean>
Main Pg 2 > System > Calibrator > Set Level
Description:
Set the output of the selected calibrator in dBm.
Menu Type:
Numeric value.
Range:
-60.0 dBm to +20.0 dbm (Internal 50 MHz)
-50.0 dBm to +20.0 dbm (External 1 GHz)
SCPI Command:
OUTPut:INTernal:LEVel:[POWer]
<numeric_value>
Main Pg 2 > System > Calibrator > Sel Calibrator
Description:
Select the active calibrator for calibration and control.
―In
t‖
selects the internal 50 MHz calibrator. ―Ex
t‖
selects an external 1 GHz calibrator. "Ext" cannot be
selected if an external calibrator is not connected.
Menu Type:
Toggle.
Choices:
Int, Ext
SCPI Command:
None.
Main Pg 2 > System > Calibrator > Status
Description:
This menu item is not implemented at this time.
SCPI Command:
None
Fig.4-57 Calibrator Menu
Internal 50 MHz
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...