Boonton 4540 Series RF Power Meter
Remote Operation
5-31
5.6.6 INITiate and ABORt Commands
The purpose of the INITiate group of commands is to start and control the process of data acquisition once a measurement
has been configured. Depending on settings, the 4540 RF Power Meter may be commanded to begin either a single
measurement (INITiate:CONTinuous OFF) which stops when complete, or enter a ―free-run
‖
mode where data acquisition
occurs continuously (INITiate:CONTinuous ON). The ABORt command terminates any operation in progress and prepares
the instrument for an INITiate command. In some operating modes, the INITiate commands do not actually start
measurements, but rather arm a hardware trigger, which is then used to gate the actual measurements cycle.
ABORt
Description:
Terminates any measurement in progress and resets the state of the trigger system. Note that
ABORt
will leave the measurement in a stopped condition with all current measurements cleared,
and forces
INITiate:CONTinuous
to OFF.
Syntax:
ABORt
Argument:
None
Valid Modes:
Any
INITiate:CONTinuous
Description:
Set or return the data acquisition mode for single or free-run measurements. If
INITiate:CONTinuous
is set to ON, the 4540 immediately begins taking measurements
(Modulated, CW and Statistical Modes), or arms its trigger and takes a measurement each time a
trigger occurs (Pulse Mode). If set to OFF, the measurement will begin (or be armed) as soon as
the INITiate command is issued, and will stop once the measurement criteria (averaging, filtering
or sample count) has been satisfied. Note that
INITiate:IMMediate
and
READ
commands are
invalid when
INITiate:CONTinuous
is set to ON; however, by convention this situation does not
result in a SCPI error.
Syntax
:
INITiate:CONTinuous <Boolean>
Argument:
<Boolean> = { 0, 1, OFF, ON }
Valid Modes:
Any
INITiate[:IMMediate[:ALL]]
Description:
Starts a single measurement cycle when
INITiate:CONTinuous
is set to OFF. In Modulated
Mode, the measurement will complete once the power has been integrated for the full FILTer
time. In Pulse Mode, enough trace sweeps must be triggered to satisfy the AVERaging setting. In
Statistical Mode, acquisition stops once the terminal condition(s) are met. In each case, no reading
will be returned until the measurement is complete. This command is not valid when
INITiate:CONTinuous
is ON, however, by convention this situation does not result in a SCPI
error.
Syntax:
INITiate[:IMMediate[:ALL]]
Argument:
None
Valid Modes:
Any
Restrictions:
INITiate:CONTinuous
must be OFF
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...