Boonton 4540 Series RF Power Meter
General Information
1-8
Table 1-2
4540 Series Performance Specifications
(continued)
(Specifications are subject to change without notice)
PULSE MODE OPERATION
Acquisition Mode:
Discontinuous triggered sample acquisition
Trace Display:
Power versus time swept trace (rolled trace for slow timebases)
Trace Averaging:
1 to 16384 samples per sweep data point, exponential
Markers (vertical cursors):
Settable in time relative to the trigger position. Marker time position is
limited to displayed trace interval.
For each marker independently:
Average, minimum and peak power at a single time offset
For a pair of markers:
Average, minimum and peak power over the interval between markers
Power ratio between markers
The following automatic measurements are selectable in the power versus time (Pulse) mode:
Pulse width
Pulse rise-time
Pulse fall-time
Pulse period
Pulse repetition frequency
Pulse duty cycle
Pulse off-time
Peak power
Pulse ―on
‖
power
dB or Percent overshoot
Waveform Average power
IEEE Top level power
IEEE Bottom level power
Edge skew (Model 4542 only)
Edge delay
MODULATED MODE OPERATION
Acquisition Mode:
Continuous (untriggered) sample acquisition
Trace Display:
Power versus time rolled trace
Signal Filtering:
―Sliding window
‖
filter; 0.002 to 16.0 seconds (fixed) or auto-filter
Channel Math:
Displays the ratio, sum (power sensors) or difference (voltage sensors)
between channels or between a channel and a reference measurement
The following automatic measurements are performed continuously in Modulated Mode:
Filtered average power
Peak power
Minimum power
Peak to Average ratio
Dynamic range
Summary of Contents for 4541
Page 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Page 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Page 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Page 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Page 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Page 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...